Toggles between single sweep and continuous sweep mode.
In continuous mode (Single (All Chans) not selected), the analyzer measures continuously, repeating the current sweep.
In single sweep mode, the measurement is stopped after the number of sweeps selected in the Define Restart dialog.
Press Single to
activate the single sweep mode and start a single sweep sequence. To start
further sweep sequences press Restart.
|
Remote control: |
Stops the current measurement and restarts a measurement sequence. In Single sweep mode a new single sweep sequence is started.
|
Remote control: |
Opens a dialog to configure the Single sweep mode.

Number of Sweeps selects the number of sweeps to be measured in single sweep mode: either one (Single Sweep) or a group of consecutive sweeps.
Scope defines whether the active sweep mode (single or continuous) and the Number of Sweeps are valid for the Active Channel only or for All Channels in the active setup. If All Channels is selected, the number of sweeps in the single sweep sequence is equal to the selected Number of Sweeps times the number of channels. The sequence starts with the first sweep in channel no. 1.
In remote control, it is possible to retrieve the results
acquired in any of the sweeps within a single sweep group; see Sweep
History.
|
Remote control: |